Common Faults And Solutions Of Desktop Ultrasonic Cleaning Machine
Oct 25, 2022
1. Water inflow is too slow
First, check whether the water inlet of the bench type ultrasonic cleaner is too slow due to the low water pressure. Check the water pressure gauge to find out the cause of low water pressure. If it is not caused by this reason, it is most likely that the water inlet filter is blocked. At this time, remove the filter and clean up the impurities.
2. Slow heating rate
There may be two reasons for this. The first one is that the heating pipe has phase loss or damage. At this time, a new heating pipe needs to be replaced. The second possibility is that the temperature sensor of the desktop ultrasonic cleaner is damaged and needs to be corrected or replaced.
3. The drainage time is too long
The drainage time is too long, which is mainly caused by two reasons. The first drainage pump does not work. At this time, it is necessary to check whether the drainage pump is short circuited and whether it is energized. If the second drain pipe is blocked, the drain pipe joint shall be removed and the impurities inside shall be cleaned.
4. The water pressure of the medical flushing tank is too low
Usually, the reason for this phenomenon is either the reverse rotation of the circulating pump of the desktop ultrasonic cleaner, or the water volume in the tank is too small. The corresponding method is to replace the lines and reinstall them, or inject more water into the tank of the desktop digital ultrasonic cleaner.
5. Unstable strength of ultrasound cleaning machine
If the inductance coil of the bechtop ultrasonic cleaner is not fixed firmly, it needs to be fixed again. If not, it usually needs to readjust the matching of the electrical inductance of the bechtop ultrasonic cleaner.
6. Large deviation between temperature display and actual value
Most of the reasons for this phenomenon are that the temperature sensor of the desktop ultrasonic cleaning machine is damaged or not connected properly. Either replace the sensor or calibrate the sensor after adjusting the connector of the temperature sensor.