Repair And Maintenance Of Ultrasonic Cleaning Machines And Correct Usage Instructions

May 09, 2023

Operation Guide
1. Check the cleaning tank to ensure that there is an appropriate amount of water inside (the water level should be at least 200mm higher than the transducer box).
2. Add an appropriate amount of cleaning agent according to the process requirements.
3. Hang the power adjustment knob to the minimum position.
4. Turn on the power switch or plug the power cable into the power outlet.
5. Turn on the power switch, turn the "power adjustment" knob to the desired power value, and the ultrasonic cleaning machine will start working.
6. When stopping work, turn the "power adjustment" knob to the minimum position and turn off the power switch.

 

Maintenance and upkeep
1. If the device is not in use for a long time, please drain the cleaning solution and clean the cleaning tank.
2. If the device is not used for a long time, please turn it on once a month according to the operating instructions and work for half an hour.
3. If there is a technical malfunction, please notify us immediately.

 

Warning
1. It is strictly prohibited to start the machine without water in the cleaning tank.
2. It is strictly prohibited to start directly with high power.
3. If the concentration of corrosive gas on site is high, please try to keep the ultrasonic system away from the cleaning tank as much as possible.
4. It is strictly prohibited to touch the electronic components inside the generator during the startup state and shutdown time of 30 minutes. The capacitor contains high-voltage electrical energy.

ultrasonic cleaner 11