Daily maintenance of ultrasonic cleaning machine

Jan 04, 2022

Common problems of ultrasonic transducer:


Ultrasonic vibrator damp, you can use megohm meter to check the plug connected with the transducer, check the insulation resistance value can judge the basic situation, the general requirements of insulation resistance is greater than 5 megohm. If the insulation resistance value is not reached, it is generally the transducer that is damp. You can put the transducer as a whole (excluding the plastic-sprayed shell) into the oven and set it to dry at about 100℃ for 3 hours or use a hair dryer to dry until the resistance value is normal.


Transducer oscillator ignition, ceramic materials broken, can be combined with the naked eye and megohm meter inspection, generally as an emergency treatment measures, can separate the damaged oscillator, will not affect the normal use of other oscillators.


Vibrator degumming, our transducer is the use of cementation, screw fastening double guarantee process, in general, this situation will not occur.


Stainless steel vibration surface perforation, the general transducer after 10 years of full load may appear vibration surface perforation.


Matters needing attention:


(1) The power supply of ultrasonic cleaning equipment and electric heater must have good grounding device.


(2) Ultrasonic cleaning equipment is strictly prohibited to boot without cleaning liquid, that is, the cleaning cylinder does not add a certain amount of cleaning liquid, shall not open the ultrasonic switch.


(3) It is strictly forbidden to open the heating switch when there is no liquid in the cleaning equipment with heating equipment.


(4) It is forbidden to use heavy objects (iron parts) to impact the bottom of the cleaning cylinder, so as not to damage the chip of the energy converter.


(5) One 220V/50Hz power supply should be used alone for ultrasonic generator power supply and equipped with voltage regulator of more than 2000W.


(6) The bottom of the cleaning cylinder should be washed regularly, and there should not be too much debris or dirt.


(7) Each time when the new liquid is changed, the parts can be washed after the ultrasonic start.