How can the ultrasonic cleaning machine maintain excellent performance?
May 07, 2022
With the more and more extensive application of ultrasonic cleaning equipment, various novel ultrasonic cleaning equipment that have been continuously improved and improved are replacing the old equipment that has been put on the market. Ultrasonic cleaning machine is developing in two aspects. First, various types of multi cylinder or transmission chain or lifting ultrasonic cleaning production lines have been put on the market one after another; The second is the development of low-frequency ultrasonic cleaning machine to high-frequency ultrasonic cleaning machine.
Of course, there may be some common problems in the use of ultrasonic cleaning machine. For example, if the ultrasonic vibrator is damp, the plug connected with the transducer can be checked with a megohmmeter. The basic situation can be judged by checking the insulation resistance value. Generally, the insulation resistance is required to be greater than 5 megohm. If the insulation resistance value cannot be reached, the transducer is generally damp. Put the transducer into the oven and set it at about 100 ℃ for drying for 3 hours, or use an electric hair dryer to remove the moisture until the resistance value is normal.
When the transducer vibrator is ignited and the ceramic material is broken, it can be checked with the naked eye and megger. Generally, it is used as an emergency treatment measure to disconnect individual damaged vibrators, which will not affect the normal use of other vibrators. For the degumming of the vibrator, our transducer adopts the double guarantee process of cementation and screw fastening, which will not happen under normal circumstances. Stainless steel vibrating surface perforation. Generally, the vibrating surface perforation may occur after the transducer is used at full load for 10 years.
In order to avoid the above problems as much as possible, the power supply of ultrasonic cleaning machine and electric heater must have good grounding devices. It is strictly prohibited to start the machine without cleaning fluid, that is, the ultrasonic switch shall not be turned on if a certain amount of cleaning fluid is not added to the cleaning cylinder; For cleaning equipment with heating equipment, it is strictly forbidden to turn on the heating switch when there is no liquid.
It is forbidden to strike the bottom of the cleaning cylinder with heavy objects (iron parts) to avoid damage to the chip of the energy converter; The power supply of ultrasonic generator shall use one 220V / 50Hz power supply separately and be equipped with voltage regulator above 2000W; The bottom of the cleaning cylinder shall be washed regularly without excessive sundries or dirt. Each time the liquid is changed, the parts can be washed only after the ultrasonic wave is started.